UltraScan VIS Spectrophotometer
Model: UltraScan VIS; Hunterlab - USA
THE CIE RANGE COLOR MEASUREMENT SPECTROPHOTOMETER
A host of measurement features and specialized sample-handling devices make UltraScan VIS a highly versatile color measurement spectrophotometer. Because of its exceptional inter-instrument agreement and long-term stability, you can be confident that differences between measurements are due to product color changes, not instrument variability. Materials on the borderline of accepted tolerances will not be unnecessarily rejected.
The UltraScan VIS easily measures both reflected and transmitted color as well as transmission haze and meets CIE, ASTM and USP guidelines for accurate color measurement. UltraScan VIS uses diffuse/8° geometry with automated specular component inclusion/exclusion. For transmission measurement, the use of a robust CIE-conforming sphere instrument (TTRAN Total Transmission mode) effectively negates the effects of minor scattering typically found in transparent samples.
The optical system has an effective bandwidth of 10 nm and spectral data is reported every 10 nm. Tristimulus color calculations are performed from 360 nm to 780 nm as recommended by the CIE. UltraScan VIS includes EasyMatch QC software.
Measurement
Measurement Principle: Dual-beam spectrophotometer
Geometry: Diffuse d/8° reflectance, d/8° total transmission, d/0° regular transmission
Spectrophotometer: Two 256 element diode arrays with a high resolution, concave holographic grating
Sphere Diameter: 152 mm (6 in.)
Sphere Coating: Spectraflect™ for sphere, Duraflect™ for port plate and specular exclusion door
Port Size/Measured Area:
+ Port Diameter/View Diameter in RSIN/RSEX reflectance modes
- Large Area View (LAV): 25 mm (1 in) illuminated/19 mm (0.75 in) measured
- Small Area View (SAV): 9.5 mm (0.375 in) illuminated/6 mm (0.25 in) measured
+ Port Diameter/View Diameter in TTRAN transmittance modes
- Large Area View (LAV): 25 mm (1 in) illuminated/17.4 mm (0.69 in) measured
- Small Area View (SAV): 25 mm (1 in) illuminated/10 mm (0.40 in) measured
+ Port Diameter/View Diameter in RTRAN transmittance mode where lens is field stop for all areas of view
- Large Area View (LAV): 17 mm (0.67 in) illuminated/17 mm (0.67 in) measured
- Small Area View (SAV): 17 mm (0.67 in) illuminated/17 mm (0.67 in) measured
Lens Switching for LAV/SAV: Automatic
Specular Component: Automated Included (RSIN) or Excluded (RSEX) in reflectance
Spectral Range: 360 nm - 780 nm full CIE visible range
Wavelength Resolution: < 2 nm
Effective Bandwidth: 10 nm equivalent triangular
Reporting Interval: 10 nm
Photometric Range: 0 to 150 %
Photometric Resolution: 0.003 % (0.01 % reported)
Light Source: Pulsed Xenon lamp, filtered to approximate D65 daylight
Automatic UV Control: 400 nm cutoff filter for UV control and UV exclusion
Optional 420 nm cutoff filter for UV exclusion
Flashes per Measurement: 1 in LAV mode (4 in. SAV mode)
Measurement Time: < 5 seconds
Transmission Modes: Total (TTRAN) and Regular (RTRAN)
Transmission Compartment: Large and open on 3 sides,
10.2 cm D X 35.6 cm W x 16.5 cm H (4 in. D x 14 in. W x 6.5 in. H)
Standards Conformance
Reflectance: CIE 15:2004, ISO 7724/1, ASTM E1164, DIN 5033, Teil 7 and JIS Z 8722 Condition C
Transmittance: CIE 15:2004, ASTM E1164, DIN 5033 Teil 7 and JIS Z 8722 Condition E, G
Haze conformance per ASTM D1003 Section 8. Procedure B Spectrophotometer
Standards Traceability: Instrument standard assignment in accordance with National Institute
of Standards and Technology (NIST) following practices described
in CIE Publication 44 and ASTM E259
Performance
Colorimetric Repeatability (20 readings range):
< 0.03 ΔE* CIE L*a*b* on white tile in LAV and SAV modes
< 0.05 ΔE* CIE L*a*b* on blue denim tile in LAV and SAV modes
Spectral Repeatability: Max 0.20 peak-to-peak between 435 nm and 695 nm
Inter-instrument Agreement:
ΔE*< 0.15 CIE L*a*b* (Avg) on BCRA II Tile Set
ΔE*< 0.25 CIE L*a*b* (Max) on BCRA II Tile Set
Physical / Electrical
Dimensions: Height: 27.9 cm (11 in.)
Width: 42.0 cm (16.5 in.)
Depth: 49.8 cm (19.6 in.)
Weight: 20.4 kg (45 lbs)
Power: 100 to 240 VAC, 47 to 63 Hz
60 watts passive, 120 watts maximum
Interface: RS-232C serial, 19,200 baud, DB9 (female) terminal
Operating Environment: 10° to 40°C (50° to 104° F), 10 % to 90 % RH, noncondensing
Storage Environment: -21° to 66°C (-5° to 150° F), 10 % to 90 % RH, noncondensing
Standard Accessories:
• Calibrated instrument white tile
• Certificate of traceability
• Black calibration light trap
• Transmittance zero calibration plate
• Green diagnostic tile
• Wavelength diagnostic filter
• Reflectance sample clamp
• LAV and SAV apertures
• RS-232C cable
• USB-to-Serial adapter
• Power cord
• EasyMatch QC Software
• EasyMatch QC Basic manual